EasyManua.ls Logo

Heidolph Hei-VAP Precision - Cleaning and Maintenance; Maintenance

Heidolph Hei-VAP Precision
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
50
Cleaning and Maintenance
Cleaning and Maintenance
Maintenance
➜ Only use genuine parts approved by the manufacturer !
The vapor tube and PTFE 26 vacuum seal must be routinely serviced.
5. Remove and clean the vapor tube and the PTFE 26 vacuum seal (see chapter
“Assembly, Glassware sets”, section “Install the vapor tube”).
6. Check the PTFE 26 vacuum seal for damages and wear, replace if necessary.
Resetting the maximum temperature limiter
In the event that the temperature controller malfunctions and the heating bath tempera-
ture exceeds the maximum temperature of 250 °C, the maximum temperature limiter
will switch off the heating bath.
✓ The heating bath is cooled and disconnect from power-supply. Power-switch off.
WARNING/
AVERTISSEMENT
Hot surfaces! / Surfaces chaudes !
Burns. / Brûlures.
➜ Cool the heating bath./
Laisser refroidir le bain chauffant.
WARNING/
AVERTISSEMENT
Verify that the basic device is only plugged to or unplugged from
the heating bath if the main switch is switched off and/or the main
power supply is disconnected./
Brancher la base de l’appareil au bain chauffant ou débrancher
la base de l’appareil du bain chauffant uniquement hors tension.
Fermer l’interrupteur.
1. Remove the heating bath liquid from the heating bath.
1
Figure 8-1: Maximum temperature limiter
2. Press the maximum temperature limiter (1), possibly with a pointed object.

Table of Contents