7
1. Unscrew the threaded tting (4) from the Woulff bottle, remove the hose clamp ring
(5).
2. Place the enclosed PTFE tube section (3) in the cap nut (2) and tighten the cap nut
(2).
3. Push the threaded connection (4) on the PTFE tube section (3), also push on the
ferrule compression tting (5).
4. Connect the Woulff bottle with the ferrule compression tting (1).
5. The vacuum valve and Woulff bottle are connected.
6. Attach the connected Woulff bottle and the vacuum valve with the mounting brackets
(6) and two screws (7) to the base unit.
✓ Suitable vacuum source is available.
✓ Suitable vacuum tube (7-8 mm) is available.
1 2 3
Figure 4-19: Vacuum connection
1 Threaded tting cap 3 Vacuum tube
2 Connector (GL hose barb)
1. Connect the vacuum hose (3) to the hose barb connector (2) to the threaded tting
cap (1).
Hose connection of individual vacuum systems, refer to the following pages.