6 ‑ Maxima XTEND AX
ERROR CODE & TYPE DESCRIPTION SOLUTION
Error 08
Electromagnetic valve error
Oil not being distributed to handpieces/
attachments
Contact Henry Schein ProRepair
Error 09
Circuit is over temperature
Circuit board malfunction Contact Henry Schein ProRepair
Error 10
Circuit detect error
Circuit board malfunction Contact Henry Schein ProRepair
SPECIFICATIONS
Power: AC 100‑240 V 50/60 Hz
Fuse: 250V 2A Slow Blow Type
Air Pressure: 58 ‑ 87 psi max (4 ‑ 6 bar)
Recommended Pressure: 65.3 psi (4.5 bar)
Air Consumption: 40 L
n
/min
Weight: 19.8 lbs (9 kg)
Dimensions: 295 (W) x 385 (H) x 295 (D)
IP Code: IP20
Classication (EN60601): Class B
Operating Temperature: 59 ~ 104°F (15° ~ 40°C)
Transport & Storage Temperature: ‑4 ~ 158F (‑20° ~ 70°C)
Operating Humidity: 25 to 90%
Transport & Storage Humidity: 5 to 95%
Operating Altitude: Below 6500 ft (2000 m)
Transport & Storage Pressure: 700 to 1060 hPa
DISPOSAL OF MEDICAL DEVICES
Any medical devices which could possibly be contaminated must rst be decontaminated by the responsible doctor or medical institution
and then be disposed of by an agent licensed and qualied to handle medical and industrial waste.
INFORMATION ON ELECTROMAGNETIC COMPATIBILITY
Guidance and manufacturer’s declaration‑electromagnetic emissions
The Maxima XTEND AX is intended for use in the electromagnetic environment specied below.
The customer or the user of the Maxima XTEND AX should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
RF emissions CISPR 11 Group 1 The Maxima XTEND AX uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B The Maxima XTEND AX is suitable for use in all establishments, including domestic
establishments, and those directly connected to the public low‑voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000‑3‑2
Class A
Voltage uctuations
/icker emissions IEC 61000‑3‑3
Compliance