Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Conducted RF IEC 61000‑4‑6
Radiated RF IEC 61000‑4‑3
3 Vrms
150 KHz to 80 MHz
3 V/m
80MHz to 2,5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the Maxima
XTEND AX including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz d = 2,3 √P 800MHz to
2,5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from xed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from
structures, objects, and people.
a
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to xed RF transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the
location in which the Maxima XTEND AX is used exceeds the applicable RF compliance level above, the Maxima XTEND AX should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re‑
orienting or relocating the Maxima XTEND AX.
b
Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.
SERVICE
For items that require service, including warranty service, please send to Henry Schein® ProRepair.
Henry Schein ProRepair
620 S. Placentia Ave.
Placentia, CA 92870 USA
Phone # 714‑579‑0175
Fax # 714‑579‑0186
WARRANTY
This product is designed for use in a dental oce and this warranty is not applicable to other uses. This product is warranted against
manufacturing defects in materials and workmanship for 2 years from date of purchase.
Henry Schein® will repair or replace product at its sole discretion. This warranty shall be limited to replacement or repair of the product
or its parts and shall not extend to any other claims, including but not limited to loss of prot, cost of removal or replacement, incidental
or consequential damages or other similar claims arising from use of this product. Damage to product resulting from acts of God, faulty
installation, misuse, tampering, accident, abuse, negligence, alterations, unauthorized repairs or problems unrelated to materials and
workmanship are not covered by this warranty.
Distributed by
HENRY SCHEIN INC.
135 DURYEA ROAD
MELVILLE, NY 11747 USA
MADE IN TAIWAN
Fabricado en Taiwan
Fabriqué à Taiwan