10/34
5 Error messages
The error message will be indicated in the speed display of the front panel
e.g. :
RPM x 100
t
START
STOP
5.1 MAINS RESET
Switch off the mains switch.
Wait at least for 10 seconds and then switch on the mains switch again.
5.2 Brief description
Display. Fault Brief description Page
- 1 - Tacho error Tacho pulses break down during the run 11
- 2 - System reset Mains interrupt 11
- 3 - Imbalance Imbalance on the motor axis 11
- 4 - Communication Communication error 12
- 5 - Overload Fault in the motor or the motor control 12
- 6 - Overvoltage Mains is out of the tolerance 13
- 7 - Overspeed Overspeed detected 13
- 8 - Undervoltage Mains is out of the tolerance 13
- 9 - Overtemperature Overtemperature switch in the motor release 14
Versions Error No indication in the speed display
Wrong machine version setting
16
- c - Controller-Watchdog Fault in power board 14
- d - Lid lock error Fault in lid lock system 15
- E - Short circuit Short circuit in power board 15
- F - No Rotor code Wrong machine version 15