9.4 Brake resistor R1
Wait at least 2 minutes after disconnecting the centrifuge from the mains, until
the intermediate circuit capacitors of the frequency converter are unloaded.
Remove the two fastening screws.
Unplug the two cable from the brake resistor.
Exchange the brake resistor.
9.5 Overtemperature fuse F3
Remove the two cables from the overtemperature fuse.
Unscrew the fastening screws.
Exchange the overtemperature fuse.
9.6 Flat ribbon cable
Unplug the flat ribbon cable from the power board.
Disassemble the control board (see section 9.3).
Unplug the flat ribbon cable from the control board.
Assemble in reverse order.
9.7 Appliance plug with mains switch
Press the upper and the lower fastening clip and remove the appliance plug out of
the housing
Remove the plugs and the earth connection from the appliance plug.
Remove the fuse holder, see chapter 7.
Exchange the appliance plug.
9.8 Imbalance switch S2
Unplug S503 from the power board.
Remove the to fastening screws.
Remove the cables at the imbalance switch.
Exchange the complete imbalance switch.
After mounting the imbalance switch, adjust it as described in chapter "Imbalance
switch-off".
9.9 Lid lock Y1
Remove the two fastening screws on the top of the upper housing.
Unplug the cable from the switch and the solenoid.
Remove the earth connection from the lid lock.
Exchange the lid lock.