9 Assembling and disassembling components
Component Abreversion Plug Connected to Description in
section
Motor M1 S401 A1 9.1
Rubber-metal bearing -- -- 9.1
Power board A1 -- -- 9.2
Control panel A3 -- -- 9.3
Flat ribbon cable -- -- A1 to A3 9.6
Lid lock Y1 S404 A1 9.9
Brake resistor R1 S405 A1 9.4
Overtemperature fuse F3 S405 A1 9.5
Mains switch Q1 -- A1 9.7
Appliance plug -- -- -- 9.7
Imbalance switch S1 S503 A1 9.8
Before assembling or disassembling components, the following working
processes must first be carried out to reach the components.
Make a note of the plug numbers !
The components are assembled in reverse order !
• Open the lid
• Switch off the mains switch
• Disconnect the centrifuge from the mains
• Remove the rotor
• Remove the fastening screws of the upper part of the centrifuge housing and place
the upper part carefully beside the centrifuge.