9 Assembling and disassembling components
Component Abreversion Plug Connected to Description in
section
Motor M1 S401 A1 9.1
Rubber-metal bearing -- -- 9.1
Power board A1 -- -- 9.2
Control panel A3 -- -- 9.3
Flat ribbon cable -- -- A1 to A3 9.6
Lid lock Y1 S404 A1 9.9
Brake resistor R1 S405 A1 9.4
Overtemperature fuse F3 S405 A1 9.5
Mains switch Q1 -- A1 9.7
Appliance plug -- -- -- 9.7
Imbalance switch S1 S503 A1 9.8
Before assembling or disassembling components, the following working
processes must first be carried out to reach the components.
Make a note of the plug numbers !
The components are assembled in reverse order !
Open the lid
Switch off the mains switch
Disconnect the centrifuge from the mains
Remove the rotor
Remove the fastening screws of the upper part of the centrifuge housing and place
the upper part carefully beside the centrifuge.