CP : control panel, FC : frequency converter, SB : supply board, CB : cooling board, CC : control cable, LL : lid locking, BC : braking chopper,
BR : brake resistor, MR : mains reset, EC : error cause, ES : error consequence, ER : error remedy, M : measurements, ECR : error-code reset
36/51
9.14. Appliance plug B4, overvoltage protection F1
Remove mains switch, see chapter 9.13.
Unscrew the two fastening screws at the appliance plug.
Pull the appliance plug out of the opening.
Remove the overvoltage protection from the appliance plug.
Replace appliance plug.
Unplug the cables at the overvoltage protection.
Replace overvoltage protection.
9.15. Imbalance switch S1
Remove plug from position X3 at the SB.
From below loosen the two fastening screws of the imbalance switch.
Loosen the fastening nuts from the motor. Lift up the motor and pick up the
imbalance switch through the opening at the centrifuge chamber.
Replace imbalance switch.
9.16. Temperature sensor B1 in centrifuge chamber
Remove plug from position X3 at the CB.
Remove the lock with appropiate tools at the front of the plug. Then press out the
four bushes.
Press out the temperature sensor in the centrifuge chamber.
Replace temperature sensor.
9.17. Temperature sensor B2 at condenser
Remove upper housing part with lid (four screws).
Lift up the upper housing part and the lid.
Unscrew temperature sensor at the mounting (one screw).
Unplug plug from position X2 at the CB.
Replace temperature sensor.
9.18. Fan M3
Remove upper housing part with lid (4 screws).
Lift up the upper housing part and the lid.
Detach fan.
Unplug the two plugs at the fan.
Replace fan.
9.19. Compressor M2
Remove upper housing part with lid (4 screws).
Lift up the upper housing part and the lid.
Loosen the four hex nuts at the compressor below.
Loosen the cable at compressor.
Replace compressor.