Preventive Maintenance
Chapter 6: General Procedures
CareAssist® and CareAssist® ES Bed Service Manual (MAN330 REV 4) Page 6 - 7
6
Preventive Maintenance Schedule
Table 6-1. Preventive Maintenance Schedule
Function Procedure
Overall condition,
frame and welded
assemblies
Check the overall condition of the bed.
Check that the structure and welded assemblies are in good working
condition and that there are no impacts or corrosion.
Do any necessary repairs or paint retouches, replace parts if
necessary.
Observe the symmetry of the bed and check that the bed frame and
base are not twisted. Do any necessary repairs or paint retouches.
Check that all labels are installed and can be read, replace them if
necessary.
Check for lose or missing hardware, replace or adjust as necessary.
Headboard and
footboard
Check the appearance, attachment and safety of the headboard and
footboard. Replace if necessary.
AC power cord Examine the plug for damage. Make sure the plug is a one-piece
molded plug assembly. If it is not, replace the plug cord assembly.
Replace any plug cord assembly that shows any of these:
Discoloration of the plug molding around the plug blades; this
could occur if the plug blades have overheated or arced.
Any signs of cracking; this could occur if the plug has been
bent and straightened to a point past its useful life.
Loose fit of the plug blade (the plug blade moves in the
molding); this could occur if the molding has overheated or
the blades have been bent and straightened to a point past their
useful life.
Replace the power cord, if damaged.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom CareAssist P1170 and is the answer not in the manual?

Hill-Rom CareAssist P1170 Specifications

General IconGeneral
ModelCareAssist P1170
TypeHospital Bed
Weight Capacity500 lbs (227 kg)
Trendelenburg PositionYes
Reverse Trendelenburg PositionYes
Battery BackupYes
Frame MaterialSteel
Knee GatchYes
Head SectionAdjustable
Foot SectionAdjustable
CPR ReleaseYes
Overall Width39 in
Height Range15 in to 30 in
Side RailsYes
Power SourceElectric
Mattress TypeFoam or Air
Patient ControlsYes
WheelsLocking Casters
WarrantyVaries