Preventive Maintenance
Chapter 6: General Procedures
Page 6 - 12 CareAssist® and CareAssist® ES Bed Service Manual (MAN330 REV 4)
SideCom®
Communication
System
Inspect and test the communication junction box.
Make sure the SideCom® Communication System features operate
correctly. Inspect the communication cable, including the male and
female pins in the plug.
Replace as necessary.
SafeView™ Alerts Make sure the Alerts operate correctly.
Repair or replace parts as necessary.
Function Procedure