22
Instrument Overview and Features
Maintain the sampling speed, even when modules are added.
Adding modules, each of which incorporates an A/D converter, does not cause the reduction
of the sampling speed.
You can perform measurement at dierent data refresh rates (sampling intervals) appropriate
for every module.
See “Measurement module data refresh intervals” in the Instruction Manual.
Setting example
• Control signals are recorded at a high sampling
interval
of 1 ms using the U8553 High
Speed Voltage Unit.
• Multi channels of temperatures are recorded at a sampling
interval
of 1 s using the U8552
Voltage/Temp Unit.
(The U8552 can use a relatively low sampling speed, allowing the lter to set a lower cuto
frequency. Using the U8552 can eliminate lower-frequency noise.)
Perform strain measurement.
You can measure multiple channels of strain data using the Strain Unit.
The instrument has a built-in bridge box, allowing it to support a variety of connection
methods.
See “Connecting a strain gage or converter” (p. 58).