54
Connecting the Cables
Connecting resistance temperature detectors
Applicable modules: U8551, LR8531
You will need: a at-head screwdriver (with a tip width of 2.6 mm) and a resistance temperature
detector
Recommended wire diameter
Single-wire Diameter of 0.32 mm to 1.29 mm (AWG 26 to 16)
Stranded wire 0.2 mm
2
to 0.52 mm
2
(AWG 24 to 20)
Standard stripping length 9 mm
1
Open the terminal block cover.
2
Press down on the terminal button with the at-head screwdriver.
3
Insert the cable into the terminal hole while depressing the button.
Three-wire type: Insert the cables into the positive terminal (red), negative terminal (black), and SoL
terminal (black).
Four-wire type: Insert the cables into the positive terminal (red), negative terminal (black), SoL
terminal (black), and SoH terminal (red).
3-wire setup
4-wire setup
Cable insulation colors vary by country and manufacturer. Check before connecting.
4
Remove the at-head screwdriver from the button.
The cable will be locked in place. Pull gently on the cable and verify that it does not come out.
5
Connect the cable to the measurement target.
6
Close the terminal block cover.
HIOKI LR8450A963-02

Table of Contents

Other manuals for Hioki MEMORY HiLOGGER LR8450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki MEMORY HiLOGGER LR8450 and is the answer not in the manual?

Hioki MEMORY HiLOGGER LR8450 Specifications

General IconGeneral
Storage MediaSD card (up to 32 GB)
Measurement TypesVoltage, Temperature, Resistance, Strain, Pulse, Logic
Sampling Rate1 ms to 1 hour
InterfaceLAN, USB, SD card slot
Power Supply100 to 240 V AC, 12 V DC
Resolution16-bit
Voltage Range±10 mV