Connecting the Cables
NOTICE
Do not use the Z2000 Humidity Sensor in locations with an excessive amount of
dust or locations causing the instrument to get wet.
Doing so could damage the Z2000 Humidity Sensor, which is not water-resistant or
dust-resistant.
The Z2000 Humidity Sensor’s sensitivity and precision will degrade over time, even under
normal operating conditions. To ensure measurement within the accuracy specications, it is
recommended to replace the sensor with a new model after one year of the rst open of the
sealed package.
If the conditions diverge from the indicated operating (storage) environment, the Z2000
Humidity Sensor’s precision may degrade in less than a year, making it incapable of accurate
measurement.
In principle, placing the Z2000 Humidity Sensor in an environment where it would be exposed to
organic gases (ketone, acetone, ethanol, toluene, etc.) may cause the sensor surface to become
contaminated, increasing humidity measurement error.
Ensure that the Z2000 Humidity Sensor is not exposed to high concentrations of chemical
solvents for extended periods of time during use or storage.
The sensor may also be contaminated by organic gases released by vinyl chloride or packaging
materials.
When not using the Z2000 Humidity Sensor, seal it along with a package of desiccant in the
polyethylene bag in which it was shipped and store it in a cool, dark place.
Exercise care to avoid condensation. The sensor is particularly prone to condensation in
environments where the temperature changes rapidly.
Proper operation is not guaranteed in the event that conditions diverge from the indicated
operating (storage) environment.
Changes in humidity (to high humidity from low humidity or to low humidity from high humidity)
will a󰀨ect measured values due to hysteresis. The Z2000 Humidity Sensor’s measured values
are a󰀨ected within a range of about 3% RH.
HIOKI LR8450A963-02

Table of Contents

Other manuals for Hioki MEMORY HiLOGGER LR8450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki MEMORY HiLOGGER LR8450 and is the answer not in the manual?

Hioki MEMORY HiLOGGER LR8450 Specifications

General IconGeneral
BrandHioki
ModelMEMORY HiLOGGER LR8450
CategoryData Loggers
LanguageEnglish

Related product manuals