1.2 Low-Pressure Gradient System
In the low-pressure gradient system, two or more solvents are
mixed in its low-pressure section from which a liquid feed is
performed by a single pump. The composition of an eluent is
made to vary with time in chromatographic separation.
Fig. 1-2 Low Gradient System
(1) Each of plural eluents is fed from the degassing device to
the low-pressure unit.
(2) In the low-pressure unit, the eluents are mixed at a pre-
specified ratio.
(3) Through the mixer, the mixture eluent is delivered.
(4) A sample is injected by the autosampler.
(5) The injected sample is separated through the column which
is maintained at a constant temperature by the column oven.
(6) Each component separated from the sample is detected as
a signal by the detector.
Primaide 1440
Fluorescence
Detector
Primaide 1310
Column Oven
Drain
Detector cell
Primaide 1210
Autosampler
Mobile phase
Wash Solution
Primaide
Organizer
Primaide 1110
Pump
Degassing
device
Mixer
Proportioning valve
Four-way joint
Column