3 - 17
3.3 GLP Functions
For this function, the following items are settable.
• Xe lamp LOGBOOK data display and resetting
• Hg lamp LOGBOOK data display and resetting
• KEY LOCK setting
• WL CHECK (wavelength accuracy check) by using Hg lamp
• CHECK (Xe lamp energy check)
NOTE: The logbook data entry is reset when a numeric value
representing “year” is input followed by pressing the
ENT key.
Even when the same year as desired is already
indicated, it must be reentered without fail.
3.3.1 Xe lamp logbook data display
Total on period of the Xe lamp and its standard lifetime (in
parentheses under Xe LAMP), the number of switching-on times
and the date of the last lamp change can be indicated.
(1) Press the
GLP (LOGBOOK=1, KEY LOCK=2,
1 WL CHECK=3, Xe CHECK=4)