3 - 8
3.2 UTLTY Setting
The UTILITY key sets the following conditions.
• OFFSET
• TIME CONSTANT
• USE PROG (whether to use time program or not)
• PMT VOLT (photomultiplier voltage)
• OTHERS
3.2.1 Offset setting
The OFFSET is a set input signal amount that is added to the
actually measured value for output.
(1) Press the
key.
(2) The following display will be presented.
<Utility initial screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_1 USE PROG=3, PMT VOLT=4, OTHERS=5)
Select OFFSET.
Press numeric key 1 and then the ENT key.
(3) The following display will be presented.
<Offset setting screen>
Set an offset value.
Press numeric key (0 to 1 0 0 0, usually input 0) and then
the ENT key.
* An offset value is settable within a range from 0 to 1000 in
increments of 1 step.
* Default value: 0