3 - 8
3.2 UTLTY Setting
The UTILITY key sets the following conditions.
OFFSET
TIME CONSTANT
USE PROG (whether to use time program or not)
PMT VOLT (photomultiplier voltage)
OTHERS
3.2.1 Offset setting
The OFFSET is a set input signal amount that is added to the
actually measured value for output.
(1) Press the
UTLTY
SET UP
key.
(2) The following display will be presented.
<Utility initial screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_1 USE PROG=3, PMT VOLT=4, OTHERS=5)
Select OFFSET.
Press numeric key 1 and then the ENT key.
(3) The following display will be presented.
<Offset setting screen>
OFFSET (0 - 1000)
_ 0
Set an offset value.
Press numeric key (0 to 1 0 0 0, usually input 0) and then
the ENT key.
* An offset value is settable within a range from 0 to 1000 in
increments of 1 step.
* Default value: 0