3 - 16
Select OTHERS.
Press numeric key 5 and then the ENT key.
(c) The following display will be presented.
<Others setting screen>
OTHERS (Em BANDWIDTH=1, ERROR OUT=2,
4 LEAK SENSOR=3, LAMP=4)
Select LAMP.
Press numeric key 4 and then the ENT key.
(d) The following display will be presented.
<Xe lamp selecting screen>
Xe LAMP ON/OFF (OFF=0, ON=1)
1
Select ON (1) or OFF (0) status of the xenon lamp.
Press numeric key (0 or 1, usually input 1) and then
the ENT key.
* Default value: 1
(e) Xenon lamp setting is now completed and the 1
st
monitor screen returns.
When the xenon lamp is lit, Xe is displayed for LAMP.
And when the lamp is extinguished, OFF is displayed
for LAMP.
<1st monitor screen>
TIME Ex Em FL PROG LAMP [I]
0.0 250 350 0.0 1 OFF
* The Xe lamp will light up when this unit is powered
on.
NOTE: A xenon lamp shortly after turning off the lights can't be
lit because of high temperature.
For re-measurement, wait at least 5 minutes or more
after turning off the lamp.