(4) Setting of first step (start conditions) of time program:
(a) Set the TIME.
<Step setting screen (TIME indicated)>
TIME Ex Em BASE PMT
_ 0.0 1
Press numeric keys 0.0 and then the ENT key.
* Be sure to enter 0.0 for TIME.
* Initial (default) value: 0.0
* The program No. currently set is indicated at the
bottom right of the screen.
The cursor now moves to Ex.
(b) Set an Ex (excitation) wavelength.
<Step setting screen (Ex indicated)>
TIME Ex Em BASE PMT
0.0 250 1
Press numeric keys 2 0 0 to 8 5 0 and then the ENT
key.
* Setting range: 200 to 850 nm
It is recommended to set the same wavelength as
was set for Ex of the measuring wavelengths.
* Initial (default) value: blank
The cursor next moves to Em.
(c) Set an Em (emission) wavelength.
<Step setting screen (Em indicated)>
TIME Ex Em BASE PMT
0.0 250 350 1