3 - 50
Select EXCHANGE STANDARD.
Press numeric key 3 and then ENT key.
(4) The following display will be presented.
<Standard Xe lamp lifetime setting screen>
XE LAMP EXCHANGE STANDARD (1 - 2500) H
_500
The standard lifetime of the Xe lamp is settable here.
Press numeric key (to input 1 to 2 5 0 0) and then the ENT
key.
* Check the part number of your Xe lamp and input the
correct value listed below.
* Default value: 500 hours
(650-1500 Xe lamp is attached with
shipment.)
(5) The setting of a standard Xe lamp lifetime is now completed
and the following display will be presented.
<Setup initial screen>
SETUP: DETECT CONTRAST S. NO.
CH1 3 02AB-111
* When you press the ENT key here, the setup items
selecting screen is presented to allow reentry.
* The selected setting will be fetched upon turning on
power supply at the next time.
(6) Turn OFF the power switch.