Annex 101
Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity tests IEC 60601
Test level
Compliance level Electromagnetic
environment –
guidance
Electrical fast
transient/burst acc.
to IEC 61000-4-4
± 2 kV for
power supply lines
± 1 kV for
input/output lines
± 2 kV for
power supply lines
± 1 kV for
input/output lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surges acc. to
IEC 61000-4-5
± 1 kV
line(s) to line(s)
± 2 kV voltage
line(s) - earth
± 1 kV
line(s) to line(s)
± 2 kV voltage
line(s) - earth
Mains power quality
should be that of a
typical commercial or
hospital environment.
Conducted RF
disturbances acc.
to IEC 61000-4-6
3 V
rms
150 kHz – 80 MHz
6 V
rms
in ISM and
amateur radio
bands between
150 kHz – 80 MHz
3 V
6 V
Portable and mobile
communications
equipment should be
used no closer to any
part of the point 2,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended safety
distance: 0.3 m
Power frequency
(50/60 Hz)
magnetic field acc.
to IEC 61000-4-8
30 A/m 30 A/m
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.