54 Operating the Device
Mode
(only point 2 AutoCPAP and point 2 BILEVEL ST20/ST25)
Selection of mode by the physician
1. Press the enter key for 1 second.
2. Use the selection keys and to select Mode.

 
3. Press the the enter key and if necessary, input the PIN code with the selection
keys and . Press the enter key after every digit.
4.
Use the selection keys and to select the desired mode. Confirm your set-
ting by pressing the enter key .
Pressure
(only point 2 CPAP and point 2 AutoCPAP and point 2 BILEVEL ST20/25 in CPAP mode)
Selection of pressure by the physician
Please note: When the pressure parameter is changed during therapy/titration, the
device adapts to the new setting slowly (25 Pa/s). This pressure ramp can be deacti-
vated using the TRENDset PC software.
point 2 CPAP:
1. Select the complete menu (see page 47).
2. Use the selection keys and to select Press..
 
  
3. Press the the enter key and if necessary, input the PIN code with the selection
keys and . Press the enter key after every digit.
4. Press the enter key and then use the selection keys and to set the desi-
red value. Confirm your setting by pressing the enter key .