60 Operating the Device
P-Max
(only point 2 AutoCPAP in APAP mode)
Selection of P-Max by the physician
1. First set P-Min.
2.
Use the selection keys and to set the desired value. Confirm your setting
by pressing the enter key .
The parameter P-Start is set next.
P-Start
(only point 2 AutoCPAP in APAP mode)
Selection of P-Start by the physician
1. First set P-Min and P-Max.
2.
Use the selection keys and to set the desired value. Confirm your setting
by pressing the enter key .
I-FLEX
(only in modes CPAP and APAP)
Inhaling under therapy pressure is always more dicult as the lungs are already partly
filled and thus also pre stressed. Such additional stress from therapy can be reduced
by means of a personalized I-FLEX setting. FLEX settings will only work if spontane-
ous respiration is actually present. FLEXLINE is not a form of artificial respiration, but
merely a way of supporting the respiratory effort of the patient.
Selection of I-FLEX by the physician
1. Select the complete menu (see page 47).
2. Use the selection keys and to select I-FLEX.
3. Press the enter key and if necessary, input the PIN code with the selection keys
and . Press the enter key after every digit.
4. Press the enter key and then use the selection keys and to set the desi-
red value. Confirm your setting by pressing the enter key .