Operating the Device 61
E-FLEX
(only in modes CPAP and APAP)
Although the point 2 is equipped with an excellent pressure control system, an E-FLEX
setting can help make expiration easier. This should however be adjusted individually
for each patient. Once again, FLEX settings will only work if actual respiratory effort
is present.
Selection of E-FLEX by the physician
1. Select the complete menu (see page 47).
2. Use the selection keys and to select E-FLEX.
3. Press the enter key and if necessary, input the PIN code with the selection keys
and . Press the enter key after every digit.
4. Press the enter key and then use the selection keys and to set the desi-
red value. Confirm your setting by pressing the enter key .