EasyManua.ls Logo

Honeywell BW RigRat - Page 72

Honeywell BW RigRat
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BW RigRat User’s Guide
72
External Ports
Serial communication, switching, and 4-20mA Input
Instrument Specifications
continued
Safe Area Charge
Port (AC CHG SAFE
AREA)
Battery charger input, DC5.7V 3A typical; Connect to AC/DC adapter.
Field charge from IS
Barrier/Solar Panel
(IS CHG
HAZARDOUS)
Line Power Supply input to extend Instrument Runtime. For Low Power version only.
RAEMet Port (Serial)
Optional; Port to RAEMet wind sensor.
Analog/Switch Input
Port
(4-20MA IN)
Optional; 4-20mA signal input, X1 channel /Switch on/off signal input, X1 channel
Relays Driver Output
(SWITCH)
Optional; Optocoupler, SPDT Normal Open, X3 channels
Languages
English, German, Spanish, Portuguese, Russian, Chinese, French, Arabic
Warranty
Two years on LEL, CO, H
2
S, and O
2
sensors
One year on other sensors
Specifications are subject to change.
Supported Sensors
Gas Type
RANGE
RESOLUTION
TEMPERATURE RANGE
COMBUSTIBLE GASES
(IR-LP)
0-100%LEL
1%LEL
-40 to +14 F /
-40 to +60° C
COMBUSTIBLE GASES
(IR-LP)
0-100%VOL
0.1%VOL
-40 to +14 F /
-40 to +60° C
COMBUSTIBLE GASES
(IR)
0-100%LEL
1%LEL
-4 to +131° F /
-20 to +50° C
COMBUSTIBLE GASES
(IR)
0-100%VOL
0.1%VOL
-4 to +131° F /
-20 to +50° C
COMBUSTIBLE GASES
(LEL)
0-100%LEL
1%LEL
-40 to +14 F /
-40 to +60° C
CARBON DIOXIDE
(CO2)
0-5%VOL/0-50000ppm
0.01%VOL/100ppm
-4 to +131° F /
-20 to +50° C