- 40 -
Step Time Temp. Oven Exhaust
1 Cal. Abt 15 min, 70 Dry Open
2 Cal. Abt. 2 min. 70 Dry Open
3 Cal. Abt. 2 min. 70 Dry Closed
4 Cal. Abt. 8 min 250 Dry Open
5 Cal. Abt. 3 min. 250 Dry Closed
6 Cal. Abt. 5 min. 190 Wet Closed
7 Cooling Abt. 30 sec 105 Open
8 Cal. Abt. 5 min. 130 Wet Closed
9 Cooling Abt. 8 min. 45 Open
10 Cal Abt. 4 min. 70 Wet Closed
ClimaOptima Calibration
Calibration should be carried out when
• CPU board is replaced
• Differential pressure sensor is replaced
• Fan wheel is removed and mounted again
or replaced
• Filter housing is replaced
The oven should be as dry as possible before calibration is carried out.
The oven temperature must not be higher than 40°C when calibration is started.
NOTE: If calibration is started by mistake, turn off the oven (no other keys are active).
NOTE: Do not interrupt the calibration process.