DURation
DURation
command/query
The DURation command allows you to specify the duration qualifier to
be used with the pattern recognizer term in generating the timing trigger.
The duration value can be specified in 10 ns increments within the
following ranges:
l
Greater than (GT) qualification
-
30 ns to 10 ms
l
Less than (LT) qualification
-
40 ns to 10 ms.
The DURation query returns the current pattern duration qualifier
specification.
Command Syntax:
:MACHine{
1
~2):lTRace:DlJRation
(GT
/
LT}, <duration-value
>
where:
GT :: = greater than
LT :: = less than
c
duration value
>
:: = real number
-
Example:
OUTPUT XXX;
“:MACHINEl:TTRACE:OLJRATION
GT.
40.OE-9”
Query Syntax:
:MACHine{
1
~P}:TTFWe:DURation?
Returned Format:
[:MACHine{
11
P):?lRace:DUFtation]
{GT 1 LT}, <duration-value>
<
NL>
Example:
10 DIM
D$[lOO]
20 OUTPUT XXX;
“:MACHINEl:TTRACE:DURATION?”
30 ENTER
XXX;D$
40 PRINT
0$
50
END