HP
54501A
PERFORMANCE
TESTS
3-5
.
BANDWIDTH
PERFORMANCE
TEST
Specification
:
Equipment
Required
:
Signal
Generator
.
. .
.
.
. .
.
.
. .
.
.
.
.
. .
.
.
. .
.
.
. .
. .
.
.
. .
.
.
. .
.
. .
.
. .
.
.
.
.
.
.
.
.
.
.
. .
.
. . .
. .
.
.
.
.
. .
.
.
.
. .
.
.
.
. .
.
.
HP
8656B
Power
Meter
.
.
. .
.
. .
.
.
.
. .
.
.
. .
.
.
.
.
. .
.
. .
.
.
.
. .
.
. .
. .
.
. .
.
.
.
. .
.
.
.
.
. .
.
.
.
.
. .
.
.
.
.
. .
.
.
.
.
. .
.
.
.
. .
.
.
.
. .
.
.
.
.
HP
436A
Power
Sensor
.
.
.
. .
.
.
. .
.
.
.
. .
.
.
.
. .
.
.
.
. .
.
.
. .
. .
.
. . . . . . .
.
.
.I .
. . . .
.
.
. .
.
.
.
.
. .
.
.
.
.
. .
.
.
.
.
.
. .
.
.
. . . .
.
. . . .
HP
8482A
Type
N
(f)
BNC
(f)
Adapter
.
.
.
.
.
. .
. . . .
.
. .
.
.
. .
.
. .
.
.
. .
.
. .
.
. .
.
.
.
.
. .
.
.
.
.
.
. .
.
.
.
. .
.
.
.
.
.
. .
.
.
.
.
. .
.
HP
1250-1474
BNC
Cable
.
.
.
. .
.
. .
.
.
.
. .
.
.
. .
.
.
.
.
. .
.
.
. .
.
.
. .
.
. .
.
. .
.
.
. .
.
.
. .
.
.
. .
.
.
. .
.
.
.
.
.
. .
. . .
. .
.
.
.
.
.
. .
.
.
.
.
. .
.
.
.
.
HP
10503A
BNC
tee
. .
.
.
.
. .
.
.
.
. .
.
.
. .
.
.
. .
.
.
.
.
. .
. . . .
.
.
. .
.
. .
.
.
. .
.
. .
.
. .
.
.
.
.
. .
.
.
.
.
.
.
.
.
. .
. . .
.
. .
.
.
.
.
. . .
. .
.
.
. .
HP
1250-0781
Procedure
:
1
.
Connectequipment
as
in
figure 3-1
.
DC
to
100
MHz
(Repetitive)
DC
to
1
MHz
(Single-shot)
2
.
Set
Signal
Generator
frequency
to
approximately
100
kHz
.
3
.
Autoscale
HP54501A
by
pressing
front-panel
AUTOSCALE
key
.
4
.
Press
front-panel
DISPLAY
key
and
select
grid
.
5
.
Press
front-panel
CHANNEL
and
set
vertical
sensitivity
to
100 mV/div
.
6
.
Adjust
Signal
Generator
output
for
vertical
deflection
on
HP
54501A
of
exactly
8
divisions
.
Adjust
offset
as
necessary
.
7
.
Set
Power
Meter
CALFACTOR
%
for
100
kHz
value
from
Power
Sensor
CAL
FACTOR
chart
.
8
.
Press
Power
Meter
%[REF]
key
to
set
0
dB
reference
.
9
.
Change
Signal
Generatorfrequency
to
100
MHz
.
10
.
Select
HP
54501A
TIMEBASE
and
set
at
10
nsec/div
.
Adjust
trigger
level,
if
necessary,
for
stable
display
.
11
.
Set
Power
Meter
CALFACTOR
%
for
100
MHz
value
from
Power
Sensor
CALFACTOR
chart
.
12
.
Adjust
Signal
Generator
output
level
to
obtain
0
dB
on
Power
Meter
.
13
.
Deflection
of
displayed
signal
should
be equal
to or
greater
than 5
.6
divisions
in
amplitude
.
14
.
Repeat
steps
1
through
14
to
test
channels
2,
3,
and
4
.

Table of Contents