HP 54501A - Delta-T Accuracy Test

HP 54501A
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HP
54501A
PERFORMANCE
TESTS
3-8
.
DELTA-T
ACCURACY
TEST
Specification
:
Delta-t
accuracy
for
dual-cursor,
single-channel
measurement,
or
for
channel-to-channel
measurement
after
visual
time
null
calibration
has
been
done
.
Equipment
Required
:
Signal
Generator
. .
.
.
. .
.
.
. .
.
.
. .
.
.
. .
.
. .
.
.
. .
.
.
. .
.
.
. .
.
. .
.
.
. . . .
.
. . . .
.
.
.
. .
.
.
.
.
. .
.
.
.
. .
.
.
.
.
.
. .
.
.
.
.
. .
.
HP
8656B
BNC
Cable
. .
.
.
.
.
. .
.
.
.
. .
.
. .
.
.
.
. .
.
. .
.
.
. .
.
. .
.
.
. .
.
.
. .
.
. .
.
.
.
.
. .
.
. .
. . .
.
. .
. . .
.
.
. .
.
.
.
. .
.
.
.
.
.
. .
.
.
.
.
. .
HP
10503A
Procedure
:
+-
2%
of
(t/div)
+-
.005%
of
Delta
t
+-
1
ns
1
.
Set Signal
Generator
for
100
MHz
output
.
2
.
Connect
Signal
Generator
to
HP
54501A
Channel
1
input
.
3
.
Autoscale
HP
54501
A
by
pressing
front-panel
AUTOSCALE
key
.
4
.
Adjust
Signal
Generator
output
level
for
6
to
8
division
vertical
deflection
on
HP
54501A
.
5
.
Change
TIMEBASE
to
5
ns/div
.
6
.
Set
TIMEBASE
delay
for 0
.00000
s
.
7
.
Go
to
DISPLAY
menu
and
set
#
of
averages
2048
.
8
.
Select
delta-t
menu
by
pressing
At/AV
key
.
9
.
Select
At
markers
on
.
10
.
Set
start
marker
to
mid-screen
crossover
point
on
left
side
of
displayed
signal
as
in
figure
3-2
.
11
.
Set
stop
marker
to
mid-screen
crossover
point
on
right
side
of
displayed
signal
as
in
figure
3-2
.
12
.
Measure
of
delta
t
at
bottom
of
display
should
be
40ns
t
1
.3
ns
(38
.7
ns
to
41
.3
ns)
.