4-18
.
Logic
Trigger
Delay
Calibration
Equipment
Required
:
BNC
Cable
. .
.
.
.
. .
.
.
.
. .
. .
.
.
. .
.
. .
.
. .
.
. .
.
.
. .
. .
.
. .
.
. . . .
.
. .
.
. . . .
.
. .
.
.
. .
.
.
. .
.
.
.
. .
.
.
. .
.
.
. . . .
.
. .
.
. . .
HP
10503A
Procedure
:
1
.
Invoke
SERVICE
MENU
by
first
pressing
front-panel
UTIL
key
and
then
selecting
service
menu
.
2
.
Select
cal
select
0
with
menu
softkey
.
3
.
Connect
BNC
cable
from
rear
panel
AC
Calibrator
Output
to
channel
1
input
.
4
.
Select
start
cal
menu
softkey
.
5
.
Select
continue
.
6
.
Adjust
System
Board
Assembly
C700
if
*
is
on
either
side of
brackets
in
message
at
bottom
of
screen
.
7
.
When
[*],
press
continue
.
8
.
When
logic
trigger
delay
cal
has
been
successfully
completed,
PASSED
message
will
be
displayed
.
9
.
Switch
rear-panel
CALIBRATION
switch
to
PROTECTED
position
.