54501A
ADJUSTMENTS
4-10
.
SYSTEM
BOARD
ASSEMBLYADJUSTMENTS
4-11
.
ExtenderCable
Installation
Equipment
Required
:
Extender
Cable
.
.
.
.
. . .
.
. .
.
.
. .
.
.
.
.
. .
.
. .
.
. .
.
. .
.
. .
.
.
. .
.
.
. . . .
. .
.
.
.
.
. .
.
.
.
.
. .
.
.
. . . .
.
.
.
. .
.
.
.
.
. .
HP
54503-61604
4-12
.
Default
Calibration
Load
No
Equipment
Required
NOTE
Use
an
external
fan
on
the
power
supply
to
prevent
the
possibility
of
ther-
mal
cut-out
of
the
power
supply
.
Procedure
:
1
.
Remove
power
from
HP
54501
A
.
2
.
Remove
top cover
.
3
.
Disconnect
power
supply cables
from
line
filter
and
System
Board
Assembly
.
4
.
Remove
locking
pins
from
power
supply
.
5
.
Slide
power
supply out
from
side
of
HP
54501A
.
6
.
Connect
the
Extender
Cable
from
power
supply
to
System
BoardAssembly
.
7
.
Reconnect
line
filter
cable
to
power
supply
.
8
.
Connect
power
cable
and
apply
power
.
1
.
Invoke
SERVICE
MENU
by
first
pressing
front-panel
UTIL
key
and
then
selecting
service
menu
.
2
.
Select
cal
select
3
with
menu
softkey
.
3
.
Set
rear
panel
CALIBRATION
switch
to
UNPROTECTED
position
.
4
.
Press
start
cal
menu
softkey
.
A
caution
message
will
be
displayed
indicating
the
cal
RAM
will
5
.
Press
continue softkey
.
7
.
Leave
rear
panel switch
in
UNPROTECTED
position
for
self
calibration
procedure
that
is
per-
formed
later
in this
section
.
be
overwritten
with
default
values
.
6
.
The
status
message
at
the
top
of
the
menu
will
indicate
default
cal
has
been
loaded
.

Table of Contents