4-13
.
DAC
Reference
Adjustment
Equipment
Required
:
Digital
Multimeter
(DVM)
.
. .
. . . .
.
. .
.
.
. .
. .
.
. . . . . . . .
.
.
. .
.
.
.
.
.
.
. .
.
.
. .
.
.
.
.
. .
. . . .
.
.
. .
.
.
. .
.
.
.
. .
.
.
.
. .
HP
3478A
Procedure
:
1
.
Invoke
SERVICE
MENU
by
first
pressing
front-panel
UTIL
key and
then
selecting
service
menu
.
2
.
Select
cal
select
2
with
menu
softkey
.
3
.
Set
dac
output
to
high
adjust
.
4
.
Connect
DVM
to
rear
panel
DC
Calibrator
Output
.
HP
54501A
ADJUSTMENTS
5
.
Adjust
System
Board
Assembly
R723
for
DVM
indication of
5
.0000
V
.
Refer
to
figure
4-4
for
adjustment
location
.
6
.
Set
cal
select
to
low view
.
7
.
Measure
V1
to
four
decimal
places,
0
.
volts
.
8
.
Set
dac
output
to
high
adjust
.
9
.
Adjust
System
Board
Assembly
R723
for
DVM
indication of
5
.000V
PLUS
V1
from
step
7
.
10
.
Select
exit
menu
softkey
and
oscilloscope
will
return to
Utility
menu
.