Display ranges
Power
-30 to -llOdBm or 1000.00mW to O.OlpW
Measurement time 20ms to lh
Calibration factor ±200.000dB
Reference ±200.000dBmj dB
Data memory 500 measurement resultsj channel
(Data acquisition time 2:20msjmeasurement result)
Environmental
Storage temperature
-40
e
C to +75
e
C
Operating temperature
o-c
to +55
e
C
Humidity
<95% R.H. from
o-c
to +40
e
C
Power
AC 100 to 240Vrms±10%, 50 to 60Hz±10%, 55VA max.
Dimensions
8.9mm H, 2l2.3mm W, 355mm D (3.5" x8.36"x14.0")
Weight
net 2.5kg (5.51bs), shipping 4.5kg (9.91bs)
HP-IB capability
modes and parameters can be programmed
Transfertime
20ms for one measurement result
(cont. trigger, watt) <800ms for 100 measurement results out of memory
HP-IB interface
Function code
SHl, AHl, T6, L4, SRl, RLl, PPO, DCl, DTl, CO, E2
Function Tests
Introduction
The procedures in this section test the electrical function of the
instrument. The complete specifìcations to which the
HP8l53A
is
tested are given in "Specifìcations".
AH
tests can be performed
without access to the interior of the instrument.
Equiprnent Required
Equipment required for the Function Test is listed in the table below.
Any equipment which satisfies the criticai specifìcations given in the
table may be substituted for recommended models.
2-2 FunctionalTests
-