installing the Analyzer
Step 2. Meet Electrical and Environmental Requirements
2. Ensure the available ac power source meets the following requirements:
1230
V
1
199
to 254
Vat
147 to 66
Hzl
1
If the ac line voltage does not fall within these ranges, an autotransformer
that provides third wire continuity to ground should be used.
3. Ensure the operating environment meets the following requirements for
safety:
l
indoor use
l
altitude up to 15,000 feet (4,572 meters)
0
temperature 0
“C
to 55
“C
l
maximum relative humidity 5 to 95 percent relative at +40
“C
(non-condensing)
l
this product is designed for use in INSTALLATION CATEGORY II and
POLLUTION DEGREE 2
CAUTION
This product is designed for use in Installation Category II and Pollution
Degree 2.
NOTE
The above requirements are for safety only Separate conditions that must be met for specified
performance are noted in Chapter 10.