Calibrating for Increased Measurement
Accuracy
Choose an Appropriate Calibration Method
How to Download
Standards Definitions
NOTE
Calibration kit definitions must be in
DOS
format.
LIF
format is not supported for cal kit definitions.
1. Insert the disk with the
ca.I
standard definitions into the
anaIyzer’s
built-in
disk drive.
2.
Press
CSAVE
RECALL) and note the first word of the second line on the
screen. If it is not
“INT:\”
(for internal disk), press
S&W%
IX&
Intsmal
3
.W
Disk Prior
Menu
.
3.
Highlight the
fle
(CALKIT,
in the case of the Example Programs disk).
4.
Press
Rerall
State and wait for the prompt:
Loaded cal kit information for 4 calibration standards
Recall of cal kit from
CALKIT
complete
5.
Press [CAL) Cal Kit
UWW
Defined.
I
NOTE
Cal kit coefficients are displayed in the cal kit block of the operating parameters screen; press
(SYSTEM OPTIONS)
Operating
Param6mxm
Mext
Smewn
.
6-18
I-
深圳市盛腾仪器仪表有限公司 Tel:0755-83589391 Fax:0755-83539691 Website: www.Sengt.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8711C and is the answer not in the manual?

HP 8711C Specifications

General IconGeneral
BrandHP
Model8711C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals