lhble
l-11.
HP-IB
Only Commands (continued)
Action Mnemonic S ? Description
OUTPUT (continued)
Limit test
OUTPSEGF’
1
Outputs the limit test status5 for a specified
3tatus
segment. See
SELSEG[D].t
Minimax OUTPSEGM’
1
Outputs limit test min/max for a specified
value
segment. See SEISEGID].t
Sequencing
OUTPSEQ<
I> 2 Outputs sequence I (I- 1 to 6) listing over
HP-R?
Serial
OUTPSERN
1
Outputs the serial number of the analyzer.
number
Status byte
OUTPSTAT
1
Outputs the status byte. ASCII format (FORM 4)
STB?
Display title
OUTPTITL
1
Outputs the display title. ASCII format (FORM
4).
Clock
READDATE
1
Outputs the date of the clock in the following
format: DD MMM YYYY
Clock
READTIME
1
Outputs the time of the clock in the following
format: HH:MM:SS
*
Refer to the “Limit Line and Data Point Special Functions” section in HPBASIC
Programming h?ramplcs Guide
5
Values returned for limit test status are: 1 (PASS), 0 (FAIL), -1 (NO-LIMIT)
I
For the definition of a limit segment, see “Example Display of Limit Lines” in the
YP
BASIC
Programming Examples G&de section titled “Limit Line and Data Point
;pecial Functions.”
HP-IB Programming and Command Reference Guide