These commands begin a calibration sequence:
CALIFUL2
Short, load, open, thru (SLOT) 2-port.
CALIONEB
One-path 2-port.
CALIBAI
Response and isolation.
CALIRESP
Response.
CALISlll Sll
l-port.
CALIS
s22 l-port.
CALITELB
Thru, reflect, line or Line, reflect, match
(TRL*/LRM*)
2-port.
These commands select a default calibration kit:
CAIX24MM 2.4-mm (HP
85056A/D
cal kit).
CALK292MM
2.92-mm.
CALK292S
2.92’ (HP
85056K
cal kit).
CALK35MD
3.5~mm
(HP
85052B/D,
HP
85033D
cal kit).
CALK35MC
3.5.mm (HP
85033C
cal kit).
Note
CAIXSBMM selects the HP
85033C
cal kit for the HP
8752C/53D.
CALKPMM
CALKN50
CALKN75
CALKTRLK
CALKUSED
7mm
(HP
85031B
cal kit and HP 85050
series).
Type-N 50 ohm (HP
85032BE
cal kit).
Type-N 75 ohm (HP
85036BE
cal kit).
TRL 3.5-mm (HP
85052C
cal kit).
I
User-defined calibration kit.
CALN
Calibration: none. Turns calibration type
to OFF.
CALPOW
Provides access to the power meter
calibration functions.
l-106
HP-IB
Progr
amming and Command Reference Guide