Begins the modify cal kit sequence.
Computes new cal set using adapter
removal.
MOD11
MODS
NEWSEQ<I>
NEXP
NOOP
NUMG[D]
NUMR[D]
Begins modifying a sequence. Range: 1 to
6.
Displays the next page of the operating
parameters fist.
No operation. OPC-compatible.
Activates D number of groups of sweeps.
A group is whatever is needed to update
the current parameter once. This function
restarts averaging if ON. OPC-compatible.
Range: 1 to 999.
Sets the number of power meter readings
per point used during a power meter
calibration. Range: 1 to 100.
OFLD
OFLS
Offset loads done.
Selects the calibration standard load
as being an offset load, as opposed to
a sliding or fixed load, during a cal kit
modification.
These 3 commands specify the offset value for the indicated parameter
for a standard being defined during a cal kit modification:
OFSD[D]
Delay offset. Range:
k
1 s.
OFSL[D]
Loss offset. Range: 0 to 2000
TWs
.
OFSZ[D]
Impedance offset. Range: 0.1 to 500
CL
OMII
Omits the isolation step of a calibration
sequence.
OPC
Operation complete. Reports the
completion of the next command received
by setting bit 0 in the event-status
register, or by replying to an interrogation
if OFT?
j
is issued.
l-130
HP-IB Programming and Command Reference Guide