Installing Your Analyzer
STEP 3. Meet Electrical and Environmental Requirements
STEP 3. Meet Electrical and Environmental
Requirements
1. Set the line-voltage selector to the
position that corresponds to the AC power
source.
2. Ensure the available AC power source
meets the following requirements:
■ 90–132 VAC
■ 47–66 Hz / 400 Hz (single phase)
OR
■ 198–265 VAC
■ 47–66 Hz (single phase)
The analyzer power consumption is 350 VA
max.
3. Ensure the operating environment meets
the following requirements:
4. Verify that the power cable is not
damaged, and that the power-source outlet
provides a protective earth contact.
■ 0 to 55 C
■ < 95% relative humidity at 40 C
(non-condensing)
■ < 15,000 feet ( ª 4,500 meters) altitude
Some HP 8753E performance parameters
are specified for 25 °C ±5 °C. Refer to the
HP 8753E Network Analyzer User's Guide
for information on the environmental
compatibility of warranted performance.
WARNING: Any interruption of the
protective (grounding) conductor or
disconnection of the protective earth
terminal, can result in personal injury, or
may damage the analyzer.