Chapter 4, Status Reporting
Status Reporting
a. Although a Condition register is implemented, the Test Set will not parse out or execute HP-IB com-
mands while this operation is taking place. Consequently, a query of the Condition register will not be
executed until after the calibration attempt has completed, and will never indicate a “true” condition.
Always query the Event register, and when the calibration attempt has completed an Event register query
result will be returned and will show which bit was latched by the initiation of a calibration attempt.
.
Table 4 Calibrating Status Register Group Condition Register Bit Assignments
Bit Number
Is Condition
Register
Implemented?
Condition Comment
15 Unused in the Test Set
14 Unused in the Test Set
13 Unused in the Test Set
12 Unused in the Test Set
11 Unused in the Test Set
10 Unused in the Test Set
9 Unused in the Test Set
8 Unused in the Test Set
7 Unused in the Test Set
6 Unused in the Test Set
5 Unused in the Test Set
4 Unused in the Test Set
3
YES
a
RF Gen Level calibration Calibrating the PCS Interface generator
power levels
2
YES
a
CDMA Baseband calibration Calibrating CDMA Baseband Physical
Channels
1YES
a
Channel power calibration Calibrating Channel Power measure-
ment.
0
YES
a
Digital power zeroing Average power measurements zeroed