62
S:\HP8924C\USRGUIDE\BOOK\chapters\opoverv.fb
Chapter 3, Operating Overview
Saving and Recalling Instrument Setups
To Recall an Instrument Setup
1. Press and release the SHIFT key, then the PREV key to access the I/O CONFIGURE
screen and select the media to recall settings from using the Save/Recall field. (The
default is internal memory.)
2. Press the RECALL key.
3. Use the knob to select the desired setup to be recalled from the Recall menu at the
bottom right of the screen.
Front-Panel Example
This example recalls the current instrument settings.
Press RECALL, 1, 2, 3, ENTER. The saved instrument settings are recalled.
HP-IB Example
The following HP-IB command also recalls register 123.
“REG:REC 123”
To Clear All SAVE Registers
1. Press the RECALL key.
2. Use the knob to position the cursor in front of the entry in the Recall menu at the
bottom right of the screen.
3. Press the knob or the ENTER key. A prompt appears at the top of the screen to verify
that you want to clear all registers.
4. Press the ON/OFF key to select YES.
HP-IB Example
The following HP-IB command clears all registers
“REG:CLE:ALL”