Chapter 3, Operating Overview
Saving and Recalling Instrument Setups
To Remove (Clear) an Individual SAVE Register
1. Specify where the register is stored using the Save/Recall field on the I/O CON-
FIGURE screen.
2. Press the RECALL key.
3. Use the knob to position the cursor in front of the register to be removed from the Re-
call menu at the bottom right of the screen. The register name and percentage of
SAVE memory occupied by that register are indicated at the very top of the screen.
4. Press the ON/OFF key. A prompt appears, asking if you want to delete the save register.
5. Press the ON/OFF key to select YES. (Press the RATIO W key to select NO.)
HP-IB Example
The following HP-IB command clears a register
“REG:CLE ‘<quoted string>’”
Choosing Register Names
You can use any number, letter, or combination of numbers and letters as a name
for storing instrument settings. For instance; if you want to save a setup for testing
a “Vulcan7” radio, you can save the setting as “VULCAN7”.
Two register names are reserved for special purposes: POWERON and BASE.
Specifying POWERON Settings
You can specify the instrument setting at power-on by following the procedure
described in
"To Save an Instrument Setup" on page 61, and choosing the register
name POWERON. If a SAVE Register named POWERON is detected by the Test
Set during its power-on routine, the Test Set will configure itself using the settings
stored in the POWERON register.
NOTE: If the Test Set does not successfully complete its power-on routine because of the POWERON
settings (e.g., the Test Set displays a message that requires you to cycle power to recover) you
must:
1. Turn off the Test Set.
2. Hold down the PRESET and the Hz/uV keys simultaneously.
3. Turn on power while holding the PRESET and the Hz/uV keys down until the CALL
CONTROL screen appears.
This procedure will clear all SAVE registers, including POWERON.