70
S:\HP8924C\USRGUIDE\BOOK\chapters\opoverv.fb
Chapter 3, Operating Overview
Interaction Between Screens
Using your Test Set, duplicate the steps in figure 7 to demonstrate how the
priority fields operate.
Figure 7 Example of How Priority Fields Work
Table 3 Priority RX TEST and TX TEST Fields
Priority Field RX TEST TX TEST
RF Gen Amplitude Presets to 80 dBm (changeable) Always Off
AFGen1 To Presets to FM (changeable) Always Audio Out
AF Anl In Always Audio In Presets to FM Demod (changeable)
Detector Always RMS Presets to Pk + Max(changeable)
De-emphasis Always Off Presets to 750µs (changeable)
AF Anl Measurement Presets to SINAD (changeable) Presets to Audio Freq (changeable)
-50
Amplitude
dBm
-50
Amplitude
dBm
-50
Amplitude
dBm
- 7 5
1. Access the RX TEST screen and set the
Amplitude to -50 dBm.
-50
Amplitude
dBm
-75
Amplitude
dBm
2. Access the RF GENERATOR screen and
set the Amplitude to -75 dBm.
3. Access the RX TEST screen. Notice that
the Amplitude changes back to -50 dBm.
2. Return to the RF GENERATOR screen.
The Amplitude is no longer where you set it
in step 2 (-75 dBm).
Since the RX TEST screen has priority control over this field, the RF GENERATOR screen’s
Amplitude setting changed when RX TEST was accessed.