EasyManuals Logo

Hydac CS1000 series User Manual

Hydac CS1000 series
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #72 background imageLoading...
Page #72 background image
Reading the analog output
SAE + T
The SAE+T signal consists of 5 measured values which are transmitted time-
coded with the following time slices:
_
Time
Signal Variable
Signal
duration
per pulse
in ms
Current (I) / Voltage (U)
1 Identifier SAE A 300 High / Low
2
Measured
value
SAE A 3000
Current/Voltage for measured value
3 Identifier SAE B 300 High / Low / High / Low
4
Measured
value
SAE B 3000
Current/Voltage for measured value
5 Identifier SAE C 300 High / Low / High / Low / High / Low
6
Measured
value
SAE C 3000
Current/Voltage for measured value
7 Identifier SAE D 300 High / Low / High / Low / High / Low / High / Low
8
Measured
value
SAE D 3000
Current/Voltage for measured value
9 Identifier Temperature 300
High / Low / High / Low / High / Low / High / Low /
High / Low
10
Measured
value
Temperature 3000
Current/Voltage for measured value
4,0 2,0
I (mA)
1
2
4
6
8
10
3
5
7
9
U (V)
time (ms)
4,8
19,2
High
High
Low
Low
300
3000
3000
3000
3000
2,25
2,4
9,6
4,5
9,7519,5
9,9
10,0
9,85
19,7
0,0
19,8
ContaminationSensor CS1000
en(us)
Page 72 / 136
BeWa CS1000 3764916d 300 en-us 2017-09-25.docx 2017-09-25