Instruction Manual IMBus
5. Calibration with 2 masters ( = Button „Calibration“ ):
- The menu „Calibration” allows the calibration of the measuring inputs with 2 masters.
- The calling of the menu is only necessary in some special cases, because the IMB-im1/2/4/8 modules
are calibrated at the manufacturer with special reference calibration equipment.
- Input passwort „scal“.
- Input the master values.
- Press the „Cal“-Button of the particular measuring input.
- After that the software requests to insert the first master and then to insert the second master. After
collection of the measuring values of both masters the software calculates automatically the gain and
offset and stores both values inside the module.
5.2 IMB-dm1/2/4
1. Short description :
- The IMB-dm1/2/4 modules allow the connection of incremental measuring systems with 1Vpp output
or together with a special adapter cable the connection of measuring systems with 11µA output.
- The pining fulfils the Heidenhain Standard.
2. Configuration :
- Open Setup-Window of IMB_Test programme.
- Press Setup-Button behind the input ( Column Gauge / Sensor ) :