M FOREWORD
The OS310 can be optionally purchased in AUV version. The AUV version foresees a reduced length
housing and special data transmission formats.
The housing length is reduced by removing the space dedicated to the internal battery pack. Therefore,
the OS310 AUV version can be powered from an external power supply only.
M.1 CONFIGURATION
It is possible to configure the OS310 to automatically transmit after the wake-up data acquired in real
time.
The sampling rate depends on the chosen rate, while data transmission format can be selected among
three different options:
a) Ocean Seven 310 custom
b) BLUEFIN-21
c) OEX-C
Ocean Seven 310 custom
Data acquired is transmitted according with the probe configuration
BLUEFIN-21
The OS310 transmit conductivity (mS/cm), temperature (°C IPTS68), pressure (dbar)
+0055.5454,+0022.3451,+0100.126
OEX-C
The OS310 transmit temperature (°C ITS90), conductivity (S/m), pressure (dbar)
0022.3451,05.55454,0022.3451,+0100.126
M.2 DATA TRANSMISSION
Data transmission can be stopped by typing <ESC> while the probe is transmitting data or SO^J if the
probe is working in NON-VERBOSE mode.