information, it is worth mentioning that the type of acquisition can be: MANUAL, TIMED, LINEAR,
CONTINUOUS, CONDITIONAL, BURST and BOTTLE. The date and time refer to the starting of the
acquisition cycle and the status can be: OK, BAD, DEL, USED.
Stored casts
------------------------------------------------------------------------------
| Data | Sensors | | EXT Mem Sector | Acquisition
Cast| Bytes |Acq. Stored| Dataset | First - Last |Stat Type Area code Date&Time
------------------------------------------------------------------------------
1 28 9 7 1 4051 4051 OK CONTIN *00112* Oct 01 12:13:04.92 2016
2 28 9 7 1 4052 4052 OK CONTIN *00112* Oct 01 12:14:07.92 2016
3 28 9 7 1 4053 4053 OK CONTIN *00112* Oct 01 12:19:00.92 2016
4 103208 9 7 3686 4054 4255 OK CONTIN *00112* Oct 01 12:37:32.92 2016
5 150976 9 7 5392 4256 4550 OK CONTIN *00112* Oct 01 12:13:12.92 2016
6 28 9 7 1 4551 4551 OK CONTIN *00112* Oct 01 12:13:40.92 2016
7 131880 9 7 4710 4552 4809 OK CONTIN *00112* Oct 01 13:52:20.92 2016
8 2100 9 7 75 4810 4814 OK TIMED *00113* Mar 28 15:03:13.92 2017
Type <any key>To continue, <ESC>To leave
The operator must select the cast to be shown from the stored ones. After choosing the cast, the following
message appears:
Show stored data records in engineering units
Press Temp Cond Sal Dens O2ppm O2sat% pH eH Time&Date
0.12 17.055 45.353 35.352 1.0258 2.35 30.1 7.895 214.8 11:14:26.8
0.19 17.010 45.304 35.350 1.0258 2.17 27.8 7.902 208.7 11:14:27.0
Type <any key>To continue
Typing a character on the PC keyboard terminates the show data function and the probe “Memory
Menu” appears again.
4.5 MEMORY DELETE DATA
This function allows stored cast data to be deleted. Deleted cast can be retrieved later-on. After invoked,
this function shows the following message:
Stored casts
------------------------------------------------------------------------------
| Data | Sensors | |EXT Mem Sector| Acquisition
Cast| Bytes |Acq. Stored| Dataset|First - Last |Stat Type Area code Date&Time
------------------------------------------------------------------------------
1 64 9 8 2 202 202 OK LINEAR ER-0112 Tue Jan 02 14:06:35.62 2007
2 640 9 8 20 203 204 OK TIMED ER-0112 Tue Jan 02 14:48:40.62 2007
3 96 9 8 3 205 206 OK MANUAL sample Wed Jan 03 12:04:38.62 2007
Select the cast to delete: 0 enter new value = 3
3 96 9 8 3 205 206 DEL MANUAL sample Wed Jan 03 12:04:38.62 2007
At the end of the deletion process, the “Memory Menu” appears again.
Delete only change the status of the cast from OK to DEL (Deleted). The status can be reverted to OK
later-on by means of the DELETE command.
4.6 INITIALIZE DATA MEMORY
By means of this function, it is possible to initialise the probe memory. Obviously, after the initialization,
all stored data will be lost. However, the operator must answer two questions before the initialization
starts. Once invoked, the following message appears:
Initialize data storage area