SECTION THREE- DATA ACQUISITION
IDRONAUT Brugherio (MB) OCEAN SEVEN 310 CTD 01-2019
32
-292.51 0.073 4.318 4.480 96.4 13.62 6.511 2.2 08:59:34.35M+
OCEAN SEVEN 310 - Timed acquisition power OFF
Present time Jun 13 08:59:34.37 2017
Next timeout Jun 13 09:00:32.37 2017
Next acquisition on Jun 13 09:00:32.37 2017
Acquisitions left to do 9
CAST temporary updating...Flush_EXT Memory...Ok done
OCEAN SEVEN 310 - ID:0317003 {USR}(1.0_10 05/17)Jun 13 08:59:34.40 2017
SHUTDOWN..RTC alarm setup..Completed
Afterwards, the probe waits in OFF condition that the next data acquisition time is met.
Note
If the probe interfaces a “Rosette” water sampling device, some questions about the water sampling method and
steps are presented to the operator. Details about the water sampling set-up can be found in the dedicated appendix.
3.13.1 Terminate a timed data acquisition
The timed data acquisition ends whenever one of the following events occurs:
Battery has run down The probe automatically checks the battery voltage against the preset
boundary each time it wakes up or at regular intervals when it is always
ON. In case the battery has run down, the following message appears:
No more power supply to continue the timed data acquisition
After this message is shown, the timed data acquisition procedure is
automatically concluded and the probe shows again the "Data
Acquisition menu". Acquired data is automatically confirmed in the
probe memory.
Operator stops the acquisition The operator can terminate the timed data acquisition by pressing the
<ESC> key while the probe is ON or by pressing <ESC> once the probe
is OFF and by answering <YES> to the question about the intention to
conclude the timed acquisition cycle. In this case, the following
message appears on the PC screen at the end of the probe start-up.
Timed data acquisition illegal acquisition time
Do you want to terminate the timed data acquisition ? (Y)es (N)o
Pressing <Y> for Yes immediately terminates the timed data acquisition.
Pressing <N> for No, the probe immediately starts an automatic power
OFF procedure calculating the next data acquisition time. The operator
has 30s to answer the question. If no answer is received from the probe,
it shuts down by itself configuring the next acquisition time.
Data acquisition error In case the probe does not complete a data acquisition procedure due
to hardware or sensor problems, the following message appears:
Error during the timed data acquisition
After this message is shown, the timed data acquisition is automatically
concluded as this error is unrecoverable. Data acquired since the
accident is automatically confirmed in memory.
Acquisition completed The timed data acquisition automatically ends whenever the total

Table of Contents