SECTION SEVEN- PROBE MAINTENANCE
IDRONAUT Brugherio (MB) OCEAN SEVEN 310 CTD 01-2019
77
Li-Ion rechargeable IDRONAUT custom battery pack 3.7, 4.5Ah
When the CTD is not used for long periods (e.g. 2 weeks or more), we suggest disconnecting the internal
battery pack connector from the CTD electronics or removing the internal battery pack from the CTD to
prevent the internal batteries from damaging the CTD due to battery acid leakage. This is why the
OCEAN SEVEN 305Plus CTD is shipped without batteries installed.
INSTALLING THREE 3.6V LITHIUM BATTERIES IN THE TRIPLE BATTERY HOLDER AND
CONNECTING IT TO THE OS310 TOP COVER WILL CAUSE A PERMANENT DAMAGE TO THE
OS310 CTD ELECTRONICS.
The status of the internal battery pack can be derived from the battery diagnostic reading of the CTD
start-up message. This last is carried out only when the laboratory cable power supply (AC/DC) is not
connected to the wall socketThree different batteries are used to operate the probe:
7.8.1 Internal battery pack endurance
The internal probe battery endurance can be easily calculated using the Os3xxCTDAutonomy” tool,
which is distributed with the ITERM program and the REDAS-5 program. This tool, through a
simplified user interface (see below), allows the operator to select the probe type, select the sensors
interfaced by the probe and then calculate the probe battery autonomy.