14 - 3
PN 074-558-P1A
3000 Micro GC Guides
15 When the analysis is complete, the chromatogram will be integrated. The areas for the peaks
identified as calibration compounds will be entered into the method. In addition, a calibration
curve will be generated using these areas. The method is now calibrated for one of the
desired levels.
16 Repeat steps 4 to 15 to calibrate other levels, one level at a time.
17 After completing a calibration in the window, click Method >> Review Calibration to review
the calibration curve and its associated data. Each peak will have its own calibration curve.
If desired, right-click on a blank area to bring up options such as View Fit Type, Change Fit
Type, and a Concentration Calculator.
Table 14-1 Conduct a multiple level calibration (continued)
Step Description