Changing a Standard Module
Table 18-1 Changing a standard module
Step Description
1 A Standard GC module must be decommissioned before it can be removed. 3000 Micro GC
internally communicates to each installed GC module via a unique module address.
For a 1-2 channel or portable instrument, the module addresses are 1 and 2.
For a 3-4 channel instrument, the module addresses are 1, 2, 3 and 4.
Decommissioning a Standard GC module will disable the GC module’s address. Open the web
browser and type the 3000 Micro GC IP Address in the address bar; press Enter to display the
embedded web page.
NOTE: At the factory, the instrument IP address is set to 10.1.1.101.
2 Click the Utilities tab.
Click Change Module Configuration.
NOTE: Two caution messages will display. Click OK on each to continue.